Pdf Summary
This Dutch registry study evaluated implantable cardioverter-defibrillator (ICD) therapy in patients with idiopathic ventricular fibrillation (IVF), a condition in which sudden cardiac arrest occurs without an identifiable cause. Among 415 patients with IVF followed for a median of 8.6 years, 32% experienced recurrent ventricular arrhythmia (VA) requiring appropriate ICD therapy, showing that arrhythmia recurrence is common and supporting the need for ICD treatment in this population. Most first recurrences occurred within 3 years, but some happened much later, even after 20 years.<br /><br />A subgroup of 261 patients who received ICDs from 2010 onward was analyzed for complications and compared between subcutaneous ICDs (S-ICD) and transvenous ICDs (TV-ICD). Inappropriate ICD therapy occurred in 13% of patients overall, with no significant difference between S-ICD and TV-ICD (13% vs 13%). The main causes of inappropriate therapy were supraventricular tachycardia, ICD malfunction, T-wave oversensing, and electrical noise. T-wave oversensing was more common in S-ICD patients.<br /><br />Device-related complications other than inappropriate therapy occurred in 11% of patients. Lead complications were the most important difference between device types and were significantly more frequent with TV-ICD than with S-ICD (13% vs 2%). Other complications such as infection, thrombosis, pocket problems, and Twiddler’s syndrome were uncommon and did not differ significantly between groups.<br /><br />Overall, the study suggests that ICD therapy is often necessary in IVF because recurrence of life-threatening arrhythmia is frequent. However, because lead complications are lower with S-ICD and inappropriate therapy is similar between systems, S-ICD may be the preferred option for IVF patients who do not need pacing.
Meta Tag
Concept
Idiopathic Ventricular Fibrillation
Concept
Implantable Cardioverter-Defibrillator
Concept
Inappropriate ICD Therapy
Concept
Lead Complication
Concept
Ventricular Arrhythmia
Keywords
idiopathic ventricular fibrillation
implantable cardioverter-defibrillator
ICD therapy
ventricular arrhythmia recurrence
sudden cardiac arrest
subcutaneous ICD
transvenous ICD
inappropriate ICD therapy
lead complications
device-related complications
Idiopathic Ventricular Fibrillation
Implantable Cardioverter-Defibrillator
Inappropriate ICD Therapy
Lead Complication
Ventricular Arrhythmia
×
Please select your language
1
English